Intralipid® Is Not Adequate for Interference Testing

Intralipid® Is Not Adequate for Interference Testing

Lipemia is best defined as turbidity in a sample caused by elevated triglycerides, mostly as chylomicrons, because they are large and more triglyceride-rich. Interference is due to a decrease in absorbance due to light scattering or, with severe lipemia, volume displacement. Turbidity is visually evident with triglycerides above about 300 mg/dL. Sample blanking and bichromatic measurements help but do not eliminate this interference.

For interference testing, lipemia is often simulated using Intralipid®, an emulsion containing 20% soybean oil, 1.2% egg yolk phospholipids, 2.25% glycerin, and water. Unfortunately, the photometric response to this synthetic “fat” differs from “physiological” lipemia. It is far more accurate to assess lipemia using intact human triglyceride-rich lipoproteins as lipoproteins are far more complex than Intralipid.

Here are the facts:

  • The nonlinear relationship between turbidity and triglyceride concentration reflects the heterogeneous nature of lipoprotein particles.
  • The L-Index derived from Intralipid does not accurately reflect triglyceride concentration.
  • The structure of synthetic liposomes in Intralipid is very different from chylomicrons.
  • Intralipid does not usually mimic interference from human triglycerides

So, why is Intralipid commonly used for interference testing: Because it is convenient. Isolating human chylomicrons from patient samples with elevated triglycerides is expensive and demanding, and requires an ultracentrifuge and other specialized equipment.

Fortunately, laboratories and diagnostic companies can now purchase concentrated human triglyceride-rich lipoproteins ready to use from Sun Diagnostics.  There are no more excuses for not performing interference testing for lipemia as accurately as possible.

References

  1. Nanji AA, Poon R, Hinberg I. Lipaemic interference: effects of lipaemic serum and intralipid. J Clin Pathol 1988; 41:1026-1027.
  2. Dimeski G. Interference testing. Clin Biochem Rev 2008; 29 Suppl (i): S43-S48.
  3. Ji JZ, Meng QH. Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assays. Clin Chim Acta 2011; 412:1550-1553.
  4. Bornhorst JA, Roberts RF, Roberts WL. Assay-specific differences in lipemic interference in native and intralipid-supplemented samples. Clin Chem 2004; 50:2197-2201.
What are the Assurance interference test kits?

Simplified Interference Testing

Sun Diagnostics simplifies interference testing with the ASSURANCE™ product line. ASSURANCE™ Interference Test Kits have been specifically designed with laboratories and manufacturers in mind.

Subscribe for News, Deals, Events, & Free Scientific Resources

The Best In Vitro Diagnostic System Designed Only For You

While we have a line of specialized products for the In Vitro Diagnostic Industry, we welcome opportunities to work with fellow scientists in conceptualizing, creating, and perfecting even more production products. Our goal is straight forward: the advancement and improvement of medicine.

Five

Product Lines

Infinite

Solutions

Custom Solutions

With multiple distribution centers and a mind at the growth of scientific study and test verification, we have custom solutions for complex scientific studies. What can we help you accomplish?