INT-01 ROUTINE INTERFERENTS
This ASSURANCE™ Interference Test Kit by Sun Diagnostics has been specifically designed with manufacturers and laboratories in mind.
ASSURANCE™ will aide manufacturers in determining assay interference and assist laboratories in verifying manufacturer’s claims for the most common interfering substances.
Human-source materials and purified chemicals are used to prepare concentrated interferents that are ready to be spiked into your specific matrix.
ASSURANCE™ Interference Test Kit for Routine Interferents PN: INT-01 Price: $1100 Includes: One vial of each component Each vial contains 1 mL unless noted |
||||
Interferent | SOURCE | ASSURANCE™ Concentration* |
Component Price | Component PN |
Triglyceride-rich lipoproteins | Human | >15,000 mg/dL | $510 | INT-01T |
Sucrose and NaCl (for TRL control ) | Pure Chemicals | 150 g/L/4.5 g/L | – | – |
Hemolysate (1.4 mL) | Human | >10,000 mg/dL | $260 | INT-01H |
Protein (1.4 mL) | Human albumin and γ-globulins | >20 g/dL | $260 | INT-01P |
Conjugated Bilirubin | Purified Chemical | >400 mg/dL | $140 | INT-01B |
Unconjugated Bilirubin | Porcine | >400 mg/dL | $140 | INT-01BU |
NaOH (for Bu control) | Pure Chemical | 0.1N | – | – |
Interferents are available individually, in customer-defined configurations, or, as INT-01. INT-01 includes one vial of each of the most common interferents.
Limitations of INT-01 ROUTINE INTERFERENTS
- Care should be used when evaluating assays for interference using native, human-sourced materials:
- Triglyceride-rich lipoprotein (TRL) material contains sucrose and sodium chloride, which may affect glucose and electrolyte assays; use appropriate control solution.
- Human albumin may bind cations (such as calcium, sodium, potassium, or magnesium), iron, thyroxine, fatty acids, bilirubin, and homocysteine.
- Drug and drug metabolites may bind to albumin, gamma-globulins, and lipoproteins.
- Typical concentrations are intended as a guide only. Laboratory should determine actual values before proceeding with interference experiment(s).
- INT-01 should not be used past the expiration date listed on the kit box. Refer to Interferent vials for expiration of individually purchased components or custom kit configurations.
- If there is evidence of microbial contamination discard the vial.
- This product is not intended for use as a control, standard or calibrator.
- Information gathered from experiments performed using ASSURANCE™ Interference Test Kit should not be used to adjust patient results. Confirm results with an alternate method when necessary.